9.  For the Mk2 Model 101 P2 Water Level Meters, cut the 

pins on the tape leads at the point show in the photo below. 

Use pliers to flatten the pins against the tape leads, so they 

fit into the terminals on the circuit board (see Step 12). 

Mk2 101 P2 Tape Connection

Pin inserted into terminal with white 

square below it on the circuit board.

10. Position the tape inside the reel hub with black plastic bridge 

over top and refasten the two screws to secure the tape to 

the reel.

11. For the Mk1 Model 101 P2, connect the Tape Molex 

Connector to the Faceplate Molex Connector.

12. For the Mk2 Model 101 P2 Water Level Meters, press 

down on the white terminals on the circuit board and insert 

the tape leads. Release the terminals and the leads should 

be secured. The lead on the bottom of the tape (numbers 

facing up) is inserted into the terminal with a white square 

below it on the circuit board. See following images.  

102

1

2

9

8

ft

7

13. With the probe in a glass of tap water, turn the Meter 

‘ON’. If the buzzer or light do not activate, check the tape 

connections. 

14. Replace the faceplate on the reel and re-secure the three 

screws.

15. Slowly wind the tape onto the reel, holding to ensure no 

slack.