Controller Board Assembly Replacement

Model 464 Mk3 250 psi

High Quality Groundwater and Surface Water Monitoring Instrumentation

(Page 1 of 2)

Tools and Materials Needed

1.  464 Mk3 250 psi Controller Board Assembly (#113871) 

 

 (Includes Circuit Board and LCD) 

2.  3/32" Allen Key
3.  3/16" Nut Driver
4.  Needle Nose Pliers

Instructions

1.  Turn the Pump Control Unit off and remove the batteries. 
2.  Use the 3/32" Allen key to undo the four hex screws on the sides 

of the Control Unit.

Hex  

Screws

Hex  

Screws

Panel

3.  Remove the panel from the Control Unit and flip it over to access 

the controller board assembly.

®Solinst is a registered trademark of Solinst Canada Ltd.

Controller Board 

Assembly

4.  To gain better access to the controller board, disconnect any 

tubing that may be in the way. To remove the tubing, push on the 

“grasping ring” of the black tubing fitting (to release tubing) and pull 

the tubing out. 

Remember where the tubing was connected

Disconnect  Tubing

Solenoid Connector Battery Connector

Solenoid Assembly 

5.  Pull to remove the battery connector and solenoid connector from 

the circuit board. 

6.  Disconnect tubing as required from the solenoid assembly. Pull to 

remove the solenoid assembly from the panel. 

Remember how it 

was connected. 

7.  Press the tabs on the label ribbon cable to unlock and pull it from 

the circuit board. You may need to use needle nose pliers.

Label Ribbon Cable

464 Mk3 250 psi Controller Board Assembly (#113871)