Levelogger User Guide - Software Version 3.4.0 or Higher
Page 26
5.3 Levelogger Status
The Levelogger Status Section of the Levelogger Settings Window shows the
Firmware Version of the attached Levelogger, an accurate battery level, the
number of free readings, the current date and time, and the start date and time
of all Leveloggers.
5.4 Setting up the Levelogger Time
The ‘Time Synchronization’ section at the middle right of the Levelogger
Settings window provides the controls for setting the Levelogger clock. The
default setting in the Enable box is off, or unchecked. If you want to synchronize
the Levelogger’s clock, check the Enable box and select the time and click the
synchronize button to set the time in the Levelogger. If you start the Levelogger
without synchronizing the clock and the time difference between the Levelogger
and the PC is more than 3 seconds, the software will give you a message
asking ‘Do you want to synchronize the clock?’. Click ‘Yes’ to synchronize the
clock. It can be very useful to synchronize the clocks of all the Leveloggers and
Barologgers going to one site.
5.5 Program Levelogger Settings
At this point you have completed editing the settings for your logging session,
click the
button to send all the settings to the Levelogger.
The Levelogger will store all the new settings. Now that the datalogger has
been programmed with your new settings, you can move on to Starting the
Levelogger.
5.6 Starting and Stopping the Levelogger
To begin logging, click the
button and the Start Levelogger dialog box will
open.
Figure 5-4 Start Levelogger Window