Levelogger User Guide - Software Version 3.4.0 or Higher
Page 29
From within the Levelogger Setting window, start the Calibration Wizard by
clicking the Calibration Wizard icon
. The wizard will guide you through
the calibration process.
You can exit the wizard after any step to cancel the calibration session.
Introduction:
The Levelogger must be connected to the computer during the
entire calibration process and
must be in stopped mode
. Ensure you use fresh
calibration solution and allow thermal equilibration of the solution during the
process. Also ensure no air bubbles are present on the sensor.
Figure 6-2 LTC Calibration Wizard Introduction
1.
Setting the Temperature Coefficient:
The default setting for the
Temperature Coefficient is 2.00. If the conductivity solution you are using to
calibrate the LTC Levelogger Junior states a different temperature coefficient
on the label, please input that number into the Temp. Coefficient field.
Figure 6-3 LTC Calibration Wizard Step 1