13.5 Restore Factory LTC Calibration
An LTC Junior Levelogger can be set back to original factory calibration
settings for conductivity at any time, using this function. If you suspect that
user calibrations are not working properly, you can restore the LTC Levelogger
Junior to its original factory settings and then perform a "first-time" conductivity
user calibration to maximize accuracy. (See Section 6).
To execute this function simply click the ‘Restore Factory LTC Calibration’
button, with the LTC Levelogger Junior connected to the PC.
13.6 LTC Calibration History Report
This function creates a report of all previous user calibrations performed on the
LTC Levelogger Junior. Use it to send a history report to Solinst for analysis
if the LTC Levelogger Junior readings are irregular and/or the unit does not
maintain its calibration.
To execute this function simply click the ‘LTC Calibration History Report’ button.
This creates a text file that can be sent to Solinst Technical Support.
Figure 13-6 LTC Levelogger Junior Diagnostics Tools
LTC Diagnostics Tools