background image

SOLOS

Automatic Lens Analyzer

Instruction For Use

Cod. 42-0001417

SOLOS

Rev. 05 – 2022

Summary of Contents for 42-0001417

Page 1: ...SOLOS Automatic Lens Analyzer Instruction For Use Cod 42 0001417 SOLOS Rev 05 2022 ...

Page 2: ...ljo na naslednjem splet nem mestu z uporabo gesla harlem https www visiaimaging com manuals php es Los manuales de instrucciones en las lenguas oficiales de los Estado miembros de la Unión Europea es tán disponibles en el siguiente sitio web usando contraseña harlem https www visiaimaging com manuals php pl Instrukcje obsługi w j ezykach urz edowych pan stw członkowskich Unii Europejskiej s a dost...

Page 3: ...op de volgende website wachtwoord gebruiken harlem https www visiaimaging com manuals php fi Euroopan unionin jäsenvaltioiden vi rallisilla kielillä olevat käyttöohjeet ovat saatavilla seuraavalla verkko sivustolla salasanan avulla harlem https www visiaimaging com manuals php no Instruksjonshåndbøker på offisielle språk i EUs land er tilgjengelige på følgende nettsted bruker passord harlem https ...

Page 4: ...al to monitor the visual defects through the measurements of spectacles and single lens optical properties 1 2 INTENDED USERS The device must be used by qualified personnel eye specialists ophthalmologists opti cians and optometrists 1 3 PLACE OF USE Health centers optician shops eye hospitals and other eye care related facilities 1 4 SERIOUS INCIDENT REPORTING In case any serious incident occurs ...

Page 5: ...t before read ing the SAFETY CAUTIONS and observe all listed instructions 2 1 DISPLAY AND SYMBOLS FOR SAFE USE Table 1 Symbol Symbols Standard Reference Description Class I Product compliant with Regulation EU 2017 745 IEC 60417 5032 Alternate current ISO 7010 M002 Read the instructions for use before usage ISO 7010 W001 Warning Sign Symbol used to alert users to potential hazards ISO 7000 3082 Ma...

Page 6: ... humidity to which the medical device can be safely exposed ISO 7000 2621 Atmospheric pressure limitation Indicates the range of atmospheric pressure to which the medical device can be safely exposed ISO 7000 0626 Keep dry Indicates a medical device that needs to be protected from moisture ISO 7000 0621 Fragile Handle with care ISO 7000 0623 This way up Indicates correct upright position of the tr...

Page 7: ...is consuming the maximum current The environmental value is stated as 40 C The device must be connected to an appropriate power source otherwise its perfor mance may be reduced Position the unit so that it is not difficult to disconnect the plug for connection to the main supply Keep this manual at hand and close to the device at all times Turn off the device if it is not going to be used for a lo...

Page 8: ...Electric shock may cause bums or a possible fire Turn the power switch OFF and unplug the power cord before replacing the fuses Replace only with fuses of the correct rating To avoid injury caused by electric shock do not open the cover Ask your dealer for service To avoid fire and electric shock in case of leakage be sure to use a grounded outlet Do not connect to outlets that are not grounded To...

Page 9: ...k is data export and remote technical assistance The SOLOS USB port must be connected to printer with USB or LAN interface Ask Topcon technical assistance for printer driver installation The SOLOS can be connected to a Local Area Network LAN through the LAN connector The network must have Ethernet protocol IEEE 802 3 Ask Topcon technical assistance and the system administrator for SOLOS and networ...

Page 10: ...ument Keep the original packaging with care in case the device needs to be transported again 2 7 CLEANING Regularly clean the device removing dust using a soft cloth In the case of more persistent surface dirt use a soft cloth soaked with water or alcohol 70 max Be careful not to wet the device and clean it only as indicated to prevent damag ing it Never use solvents or other abrasive agents ...

Page 11: ...al packaging for future use The system must always be handled shipped in its original packaging which is specifically designed to protect it against damages Before installing the system read chapter 2 in this manual The steps are illustrated in the following pictures 1 2 3 4 5 11 ...

Page 12: ...4 A CC E S S O R I E S Check Calibration Tool Power cable country specific Instruction for Use Printer Paper Cartridge ...

Page 13: ...0 01 0 0625 0 125 0 25 D Cylinder axis from 0 to 180 step 1 Addition power from 4 D to 4 D step 0 01 0 0625 0 125 0 25 D Prism power from 0 D to 20 D step 0 01 0 0625 0 125 0 25 D Prism base from 0 to 360 step 1 PD Measurement Mono Bino Spectrometer range from 315 nm to 800 nm Electrical specifications Power supply AC 100 240 V 50 60 Hz Power input 50 VA Fuse 2 5 A L 250 V dimensions 20x5 mm Envir...

Page 14: ...Linux Custom Image Processor Icore MX8 Mini RAM 1 GB Hard Disk 4 GB eMMC External connections 2 USB 1 Ethernet 1 RS 232 Connectivity 802 11 b g n Wifi LAN Mechanical specifications Width 245 mm Height 437 mm Depth 354 mm Net Weight 11 kg ...

Page 15: ...d for non interruption by fuses In this case you do not identify risks to the patient but on the user or damage to the device and then stop working and possible data loss Remove the blown fuse from its seat and replace it with an identical as indicated in tech nical specifications Make sure that the power switch of the main body is off and the power cable is unplugged The procedure to change the f...

Page 16: ... 6 7 Figure 1 Fuse changing procedure ...

Page 17: ...7 O P E R AT I N G A N D U S A G E M E T H O D Download the user manual from the following url using password harlem https www visiaimaging com manuals php 17 ...

Reviews: