6
M A I N T E N A N C E
No special scheduled maintenance is necessary for the product.
6.1 CHANGING THE FUSES OF SOLOS DEVICE
Make sure that the power switch of the main body is off and the power cable is
unplugged.
It is mandatory to use fuses only with the indicated characteristics.
The use of undersized fuses can cause the interruption of power to the device during
normal working conditions. In this case there is no risk to the user, nor for the patient,
but the device turns off at inopportune moments, and this can cause data loss.
The use of oversized fuses can lead to damage to the internal electronics of the device
due to current overload for non-interruption by fuses. In this case you do not identify
risks to the patient, but on the user or damage to the device and then stop working and
possible data loss.
Remove the blown fuse from its seat and replace it with an identical as indicated in tech-
nical specifications.
Make sure that the power switch of the main body is off and the power cable is
unplugged.
The procedure to change the fuse is the following (see
1
open the fuse box cover using a screwdriver;
2
take out the fuse box (use a screwdriver to release it);
3
substitute the fuses and make sure they stay in the correct position;
4
push the fuse box carefully back into position;
5
close the black cover again and check if correct voltage is displayed in the opening.
Table 3:
Fuse Types
Power supply voltage
Fuse type
Fuse value
220
-
240 V
20
x
5 mm
T
2
.
5 A
L antisurge