The ORS-2000 Series Solution Warming System is designed to provide controlled 
warming for irrigation solutions. It is not intended as a blood warmer. 

 

Any reference to a model includes all versions of that model (eg. ORS-2066 includes 
ORS-2066 and ORS-2066D and ORS-2066R). 

 

IMPORTANT LABELS AND INDICATORS 

 

 

Temperature Control 

 
 

Power - When lit indicates unit is on 

 
 

Serial Number & Ratings 

 

DANGER:  RISK  OF  EXPLOSION  IF  USED  IN  THE  PRESENCE  OF 

FLAMMABLE ANESTHETICS.

 

 

CAUTION: FOR THE ORS-2038 AND ORS-2058 WARMING MACHINES: DO 

NOT OPERATE WARMING MACHINE WITHOUT A MINIMUM OF 
500 mL SOLUTION IN THE ORS APPROVED DRAPE. 

 
 

FOR  THE  ORS-2057,  ORS-2066  AND  ORS-2094  WARMING 
MACHINES: 

DO 

NOT 

OPERATE 

WARMING 

MACHINE 

WITHOUT  A  MINIMUM  OF  750  mL  SOLUTION  IN  THE  ORS 
APPROVED DRAPE. 

 
 

FOR  THE  ORS-2075  WARMING  MACHINE:  DO  NOT  OPERATE 
WARMING  MACHINE  WITHOUT  A  MINIMUM  OF  1500  mL  
SOLUTION IN THE ORS APPROVED DRAPE. 

 
CAUTION: ELECTRIC  SHOCK  HAZARD,  DO  NOT  REMOVE  PROTECTIVE 

PANEL. 

REFER 

SERVICING 

TO 

QUALIFIED 

SERVICE 

PERSONNEL. 

 
WARNING: RISK OF FIRE. REPLACE FUSE AS MARKED: 
 

250V 3A TIME DELAY. 

Summary of Contents for ORS-2000 Series

Page 1: ...OPERATORS MANUAL for the ORS 2000 Series Irrigation Solution Warming System O R Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 343 6771 703 834 0101 703 834 0966 Fax 3118 B...

Page 2: ...N IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS CAUTION FOR THE ORS 2038 AND ORS 2058 WARMING MACHINES DO NOT OPERATE WARMING MACHINE WITHOUT A MINIMUM OF 500 mL SOLUTION IN THE ORS APPROVED DRAPE...

Page 3: ...1 Upper indicator PV Present value indicates present temperature of solution 2 Lower indicator SV Set value indicates the temperature the solution is set to warm to 3 AL lights when Present Value is 4...

Page 4: ...perature to desired setting using up down arrows Step 3 Press enter Step 4 Press index F Depress POWER switch to turn machine off Do not turn off by pulling power cord from outlet G Discard solution a...

Page 5: ...emperature Controller verification The ORS 2000 Series Warmer temperature controller should be verified every 5 years O R Solutions Inc requires the use of the following items or equivalent fluid temp...

Page 6: ...r replace defective warmers Improper use and or treatment of the warmer voids this warranty at the option of O R Solutions Inc O R SOLUTIONS INC DISCLAIMS ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INC...

Page 7: ...ch set point temperature O R Solution Make sure that the temperature is adjusted properly see page 3 If the temperature adjustment is correct inspect the drape The drape should be evenly distributed a...

Page 8: ...120o F 49 o C Temperature Variance 2o F 1 o C Temperature Control Microprocessor temperature controller LED readout of temperature and Set Point temperature Model Capacity Bowl Dimensions ORS 2038 3...

Page 9: ...UNTING INSTRUCTIONS 1 Assembling ORS 2000 Series Solution Warmer with the ORS S21 or ORS S22 Stand 2 Mounting of ORS 2038 Solution Warmer to ORS 1075HS Slush Machine Mounting Screws 2x ORS 2038 Irriga...