Page 2 of 9
3118.B
OPERATORS MANUAL
INTRODUCTION
The ORS-2000 Series Solution Warming System is designed to provide controlled
warming for irrigation solutions. It is not intended as a blood warmer.
Any reference to a model includes all versions of that model (eg. ORS-2066 includes
ORS-2066 and ORS-2066D and ORS-2066R).
IMPORTANT LABELS AND INDICATORS
Temperature Control
Power - When lit indicates unit is on
Serial Number & Ratings
DANGER: RISK OF EXPLOSION IF USED IN THE PRESENCE OF
FLAMMABLE ANESTHETICS.
CAUTION: FOR THE ORS-2038 AND ORS-2058 WARMING MACHINES: DO
NOT OPERATE WARMING MACHINE WITHOUT A MINIMUM OF
500 mL SOLUTION IN THE ORS APPROVED DRAPE.
FOR THE ORS-2057, ORS-2066 AND ORS-2094 WARMING
MACHINES:
DO
NOT
OPERATE
WARMING
MACHINE
WITHOUT A MINIMUM OF 750 mL SOLUTION IN THE ORS
APPROVED DRAPE.
FOR THE ORS-2075 WARMING MACHINE: DO NOT OPERATE
WARMING MACHINE WITHOUT A MINIMUM OF 1500 mL
SOLUTION IN THE ORS APPROVED DRAPE.
CAUTION: ELECTRIC SHOCK HAZARD, DO NOT REMOVE PROTECTIVE
PANEL.
REFER
SERVICING
TO
QUALIFIED
SERVICE
PERSONNEL.
WARNING: RISK OF FIRE. REPLACE FUSE AS MARKED:
250V 3A TIME DELAY.