The data acquired on multiple runs is stored in a single SIF data file. Each channel of
acquired data is tagged with the designated run number.
Run descriptions are stored in a fixed 2048 character record of the SIF data file
allocated when the test is initialized. The run descriptions can be extracted from the
SIF data file using an EASE procedure file that is resident in the TCE installation
directory.
4.3.5
Using Interactive Triggers
TCE interactive triggers allow user input from the software to control a computed
channel or DataMode through the use of an Interactive Trigger computed channel.
There are eight Boolean logic controls available which must be assigned to an
Interactive Trigger computed channel. When the test is initialized, TCE sets all the
triggers to FALSE. Select Interactive Triggering from the Test Control menu to control
the trigger states before or during a test run. Use the check box by each trigger index
and clack Apply Triggers Now to change the trigger state. A check indicates a TRUE
state. TCE grays out the triggers not defined in the test setup. To invert the trigger logic
(i.e., a check for a FALSE state), use the invert trigger parameter in the Interactive
Trigger computed channel. For more information on the Interactive Trigger computed
channel, see
“Interactive Trigger” on page 145
.
NOTE
The eDAQ
lite
maintains the trigger states through power fail or error reset test run
restarts.
4.3.6
Stopping a Test Run
Manually stop the current test run by selecting Stop Run from the Test Control menu
or toolbar or use the Test Run Stopper computed channel to automatically stop a test
run based on triggering conditions. For SIF files, certain post run tasks may be
required before the test is actually stopped. If the PCM storage option is in use, the
SIF post run tasks consist of ensuring that the SIF component files on the PC Card
are completely flushed and closed. There are no post run tasks for SIE data files.
4.3.7
Ending a Test
To end a test run, select End Test from the Test Control menu or toolbar. Ending a test
prevents starting a new test run until after re-initializing the eDAQ
lite
. This option is
available when a test is initialized and no test run is in progress. Data files are still
available for upload after ending a test.
4.4
Monitoring Test Status
Select Get Test Status from the Test Control menu or toolbar to display the state of the
current test and eDAQ
lite
data storage. The provided indicators are described below.
HBM: public
Summary of Contents for eDAQlite
Page 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Page 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Page 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Page 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Page 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Page 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Page 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Page 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Page 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Page 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Page 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...