CMM-30 – USER MANUAL
70
11.2 Operating data
a)
measurement category according to IEC 61010-1 ...................... CAT IV 600 V (III 1000 V)
b)
type of insulation ........................................................................................... double, Class II
c)
degree of housing protection acc. to EN 60529 ............................................................. IP67
d)
pollution degree .................................................................................................................... 2
e)
power supply of the meter ................................................................... 4 x AAA 1.5 V battery
............................................................................ 4 x AAA NiMH 1.2 V rechargeable battery
f)
diode test ........................................................................................... I = 1 mA, U
0
< 3 V DC
g)
continuity test ............................................................................ acoustic signal for R < 30 Ω
h)
indication for range exceeding .............................................................................OL symbol
i)
low battery indication .......................................................................................... symbol
j)
measurement rate ...................................................................3 measurements per second
k)
response time for MIN/MAX function ........................................................................ 100 ms
l)
response time for PEAK function .................................................................................. 1 ms
m)
input impedance ........................................................................................ 10 MΩ (V AC/DC)
n)
input impedance for Low Z ............................................................................... approx. 3 kΩ
o)
AC read-out ...............................................................................True RMS (A AC and V AC)
p)
AC bandwidth .................................................................................................... 45…1000 Hz
q)
LCD display .............................................................................................. 4 digit with backlit
...................................................................................... 6000 counts with function indicators
r)
dimensions ................................................................................................. 170 x 75 x 48 mm
s)
meter weight .................................................................................................................. 418 g
t)
meter weight (without batteries) .................................................................................... 371 g
u)
fuses........................................................................... mA and µA range: 0.8 A / 1000 V fast
.................................................................................................... A range: 10 A / 1000 V fast
v)
operating temperature ............................................................................................. 0..+40
C
w)
operating humidity ........................................................................................................... 75%
x)
storage temperature ............................................................................................. -20..+60
C
y)
storage humidity ........................................................................................................... < 80%
z)
maximum operating altitude ....................................................................................... 2000 m
aa)
Auto-Off function ............................................................................................ approx. 15 min
bb)
compliance with the requirements of the following standards ...............................................
.............................................................................................................................EN 61010-1
........................................................................................... EN 61010-031, EN 61010-2-033
....................................................................................................EN 61326-1, EN 61326-2-2
cc)
quality standard ...................................................................................................... ISO 9001
Summary of Contents for CMM-30
Page 1: ...INSTRUKCJA OBS UGI 1 USER MANUAL 38 MANUAL DE USO 75 CMM 30 v1 03 26 08 2019...
Page 2: ......
Page 3: ...1 INSTRUKCJA OBS UGI MULTIMETR PRZEMYS OWY CMM 30 Wersja 1 03 26 08 2019...
Page 12: ...CMM 30 INSTRUKCJA OBS UGI 10 4 Opis funkcjonalny 4 1 Gniazda i funkcje pomiarowe...
Page 40: ...38 USER MANUAL INDUSTRIAL MULTIMETER CMM 30 Version 1 03 26 08 2019...
Page 49: ...CMM 30 USER MANUAL 47 4 Functional Description 4 1 Measuring terminals and functions...
Page 77: ...75 MANUAL DE USO MULT METRO INDUSTRIAL CMM 30 Versi n 1 03 26 08 2019...
Page 86: ...CMM 30 MANUAL DE USO 84 4 Descripci n funcional 4 1 Tomas y funciones de medici n...
Page 112: ...CMM 30 MANUAL DE USO 110 NOTATKI NOTES NOTAS...
Page 113: ...CMM 30 MANUAL DE USO 111 NOTATKI NOTES NOTAS...
Page 114: ...CMM 30 INSTRUKCJA OBS UGI 112 NOTATKI NOTES NOTAS...
Page 115: ......
Page 116: ...CMM 30 INSTRUKCJA OBS UGI 2...