CMM-40
– USER MANUAL
67
10 Storage
In the case of storage of the device, the following recommenda-
tions must be observed:
Disconnect all the test leads from the meter,
Make sure the meter and its accessories are dry,
In the case the meter is to be stored for a prolonged period of
time, the battery must be removed from the device,
11 Dismantling and utilization
Worn-out electric and electronic equipment should be gathered
selectively, i.e. it must not be placed with waste of another kind.
Worn-out electronic equipment should be sent to a collection
point in accordance with the law of worn-out electric and electronic
equipment.
Before the equipment is sent to a collection point, do not dis-
mantle any elements.
Observe the local regulations concerning disposal of packages,
worn-out batteries and accumulators.
12 Attachments
12.1 Technical data
“m.v.” means measured value of standard.
DC voltage measurement
Range
Resolution
Basic uncertainty
400,00mV
0,01mV
(0.06% m.v. + 4 digits)
4,0000V
0,0001V
40,000V
0,001V
400,00V
0,01V
1000,0V
0,1V
(0.1% m.v. + 5 digits)