64
8
Replacing the battery
CMM-60 is powered from a battery pack of 7.4 V 2400 mAh.
Note:
While performing the measurements with the battery discharged, the user must be aware of
additional unspecified uncertainties or unstable operation of the device.
WARNING:
If the test leads are left in the terminals during replacement of the battery, they may generate a
risk of electric shock.
To replace the battery:
remove test leads from the terminals of the meter.
set the rotational function selector at OFF,
deflect the leg and unscrew 4 screws fixing the compartment cover,
remove the cover,
remove the battery pack and insert a new one, observing the polarity,
put on the cover and tighten the fixing screws.
WARNING:
To avoid electric shock, do not use the meter if the battery compartment cover is not in place
or is not properly fastened.
Note:
If the meter does not work properly, check the fuses and the battery pack in order to ensure
that they are in proper condition and properly installed in the device.
Li-Pol battery
pack 7.4 V
Battery cover
Leg