CMP-1015-PV – USER MANUAL
63
6.5 INRUSH function
The INRUSH function captures the starting current precisely in
the beginning of 100-millisecond period when the device is just
started. To perform the measurement:
activate the AC measurement,
press and hold
MENU/INRUSH
button,
fasten the clamp on the cord supplying power to the tested object,
turn ON the tested object,
read the results.
6.6 MAX/MIN function
To activate the mode, use the buttons
to select the
MAX
option.
Max
symbol
– the meter displays the highest value out of ex-
isting measurement readings.
Min
symbol
– the meter displays the lowest value out of exist-
ing measurement readings.
To deactivate the function, select the
MAX
option again.
When the function is activated, the automatic adjustment
of the measuring range is not available.
If the reading is outside the measurement range, symbol
OL
is displayed.
6.7
button
Briefly press
, to turn on or off the flashlight mode
6.8 MENU button and main menu
Press the
MENU
button shortly to call up the main menu.
Use the ▲▼ buttons to select the item.
Use the ► or
MENU
button to enter the desired item.
Use the ◄ button to return to the higher level menu.
To exit the main menu, use the
◄
,
MODE/VFD
,
RANGE
,
HOLD/REL
. buttons.
Summary of Contents for CMP-1015-PV
Page 1: ...INSTRUKCJA OBS UGI 1 USER MANUAL 39 MANUAL DE USO 77 113 CMP 1015 PV v1 00 14 03 2022...
Page 2: ......
Page 12: ...CMP 1015 PV INSTRUKCJA OBS UGI 10 4 Opis funkcjonalny 4 1 Gniazda i funkcje pomiarowe...
Page 50: ...CMP 1015 PV USER MANUAL 48 4 Functional Description 4 1 Measuring terminals and functions...
Page 88: ...CMP 1015 PV MANUAL DE USO 86 4 Descripci n funcional 4 1 Tomas y funciones de medici n...
Page 115: ...113 CMP 1015 PV 1 00 14 03 2022...
Page 119: ...CMP 1015 PV 117 1 Sonel CMP 1015 PV CMP 1015 PV CMP 1015 PV...
Page 120: ...CMP 1015 PV 118 2 2 1 PN EN 61010 1 2010 AMD1 2016 60 35 30 RMS 16 RMS 42 4 22 6 OL...
Page 121: ...CMP 1015 PV 119 A DC A AC 1000 A V DC V AC 1500 1000 RMS 1000 RMS 1000 2 2 II...
Page 122: ...CMP 1015 PV 120 3 COM OFF 15 60 OFF...
Page 123: ...CMP 1015 PV 121...
Page 124: ...CMP 1015 PV 122 4 4 1...
Page 125: ...CMP 1015 PV 123 MODE VFD o o LoZ o VFD MENU INRUSH o o o o o HOLD REL o o HOLD o REL...
Page 127: ...CMP 1015 PV 125 4 2 V A AC DC 30 VFD VFD HVDC F n m k M OL Auto Range H HOLD LoZ MAX MIN...
Page 128: ...CMP 1015 PV 126 4 3...
Page 129: ...CMP 1015 PV 127 5 5 1 60A 600A 1000 A MODE VFD...
Page 130: ...CMP 1015 PV 128 HOLD REL 5 2...
Page 131: ...CMP 1015 PV 129 5 3 0 V AC DC MODE VFD AC DC AC DC COM V LoZV CAP Hz K Temp...
Page 133: ...CMP 1015 PV 131 5 7 CAP COM V LoZV CAP Hz K Temp 5 8 CAP COM V LoZV CAP Hz K Temp MODE 50...
Page 134: ...CMP 1015 PV 132 5 9 CAP COM V LoZV CAP Hz K Temp MODE V 0 7 0 3 V 1 2 5 0 OL 0...
Page 135: ...CMP 1015 PV 133 5 10 CAP MODE nF COM V LoZV CAP Hz K Temp...
Page 136: ...CMP 1015 PV 134 5 11 K Temp C F MODE COM V LoZV CAP Hz K Temp K...
Page 142: ...CMP 1015 PV 140 7 CMP 1015 PV Li Pol 7 4 1200 OFF...
Page 143: ...CMP 1015 PV 141 8 7 8 9 10 11 12 60...
Page 144: ...CMP 1015 PV 142 9 10...
Page 151: ......
Page 152: ...CMP 1015 PV 2...