35
AC Voltage (True RMS)
Range
Resolution
Basic uncertainty
Input
impedance
6,6V
0,001V
10M
Ω
66V
0,01V
660V
0,1V
750V
1V
±(1,5% m.v. + 8d)
for 50...500Hz
9,1M
Ω
Crest Factor:
≤
3.
AC coupled true rms specified from 5% to 100% of range.
Frequency ranges: 50Hz ~ 1kHz. Accuracy: ±(0.1% m.v. + 5 d). The
readings of frequency measurements will show on the sub-display.
Minimum Input voltage Range: >500dgts.
Overload protection: 1000V DC or 750V AC rms.
AC current (True RMS)
Range Resolution
Basic uncertainty
±(2,0% m.v. + 10d) for 50...60Hz
600A
0,1A
±(3,0% m.v. + 10d) for 61...400Hz
±(2,5% m.v. + 10d) for 50...60Hz
and 660...1000A
±(3,5% m.v. + 10d) for 61...400Hz
and 660...1000A
1500A
1A
±(5,0% m.v. + 10d) for 50...400Hz
and 1000...1500A
Crest Factor:
≤
3.
AC coupled true rms specified from 5% to 100% of range.
Frequency ranges: 50Hz ~ 1kHz. Accuracy: ±(0.1% m.v. + 5 d). The
readings of frequency measurements will show on the sub-display.
Minimum Input current Range: >500dgts.
Overload protection: 1500A AC.