CMP-400 – USER MANUAL
42
5.4
Continuity Measurements
WARNING:
Measurements must not be realised in live circuits. Capaci-
tors must be discharged.
WARNING:
Do not realise measurements if the battery compartment is
open.
In order to realise continuity test it is necessary to realise the
following actions:
place the rotational selector in the position
Ω
,
connect the red test lead to sockets
V/
Ω/TEMP
and the black
one to sockets
COM
,
press the
MODE
button until
appears in the display,
make sure symbol
is displayed with open test leads, and in
the case of shorted terminals of the measurement probes the
dis
play shows „0”,
touch the test probe tips to the circuit under test,
read the result of the measurement on the display; if the re-
sistance is < 50
Ω, a tone will sound,
having done the test disconnect the test leads from the meter.
5.5
Diode Measurements
WARNING:
Measurements must not be realised in live circuits. Capaci-
tors must be discharged.