18
13 Technical data
•
The „m.v.” means the measured value of standard.
Type K temperature range
IR measurement
IR temperatu-
re range
D:S
Resol
ution
Basic accuracy
-32…-20
℃
-25,6…-4
℉
±5
℃
±9
℉
-20…200
℃
-4…392
℉
±(1,5% m.v + 2
℃)
±(1,5% m.v + 3,6
℉
)
-32…380
℃
-25,6…716
℉
13:1
0,1
℃
0,1
℉
200…380
℃
392...716
℉
±(2,0% m.v + 2
℃)
±(2,0%m.v + 3,6
℉
)
Other technical data
a) response time ….………..……….……..…… less than 1 second
b) spectral response ...………….………...………………….8~14
µ
m
c) emissivity………………..…...digitally adjustable from 0,10 to 1,0
d) over range indication……..…………… LCD will show “-0L”, ”0L”
e) polarity...………….automatic (no indication for positive polarity);
minus (-) sign for negative polarity
f) diode laser……………output <1mW,Wavelength 630~670nm,
class 2 laser product
Temperature
range
Resolution
Basic accuracy
-50,0…999,9
℃
-58,0…999,9
℉
0,1
℃
0,1
℉
±(1,5% m.v. + 3
℃)
±(1,5% m.v. + 5
℉
)
1000…1370
℃
1000…2498
℉
1
℃
1
℉
±(1,5% m.v. + 2
℃
)
±(1,5% m.v.+ 3,6
℉
)