DIT-500
– USER MANUAL
35
Emissivity values
Substance
Thermal
emissivity
Substance
Thermal
emissivity
Asphalt
0,90 to 0,98
Cloth (black)
0,98
Concrete
0,94
Human skin
0,98
Cement
0,96
Lather
0,75 to 0,80
Sand
0,90
Charcoal
(powder)
0,96
Earth
0,92 to 0,96
Lacquer
0,80 to 0,95
Water
0,92 to 0,96
Lacquer (matt)
0,97
Ice
0,96 to 0,98
Rubber (black)
0,94
Snow
0,83
Plastic
0,85 to 0,95
Glass
0,90 to 0,95
Timber
0,90
Ceramic
0,90 to 0,94
Paper
0,70 to 0,94
Marble
0,94
Chromium oxides
0,81
Plaster
0,80 to 0,90
Copper oxides
0,78
Mortar
0,89 to 0,91
Iron oxides
0,78 to 0,82
Brick
0,93 to 0,96
Textiles
0,90
5 Measurement operation
Hold the meter by its handle grip and point it toward the sur-
face to be measured.
Pull and hold the trigger to turn the meter on and begin testing.
The display will light if the battery is good. Replace the battery
if the display does not light.
Release the trigger and the HOLD display icon will appear on
the LCD indicating that the reading is being held. In HOLD sta-
tus, press the
①
button to turn on/off the laser and turn on/off
backlight.
The meter will automatically power down after approximately 7
seconds after the trigger is released (unless the unit is locked
on).
Summary of Contents for DIT-500
Page 1: ...1 INSTRUKCJA OBS UGI 1 USER MANUAL 25 MANUAL DE USO 47 DIT 500 v1 7 05 03 2020...
Page 2: ......
Page 27: ...25 USER MANUAL NON CONTACT INFRARED THERMOMETER DIT 500 Version 1 7 05 03 2020...
Page 71: ......
Page 72: ......