KT-160 USER MANUAL
102
Film Path: Sets the directory where the video files will be stored (default subdirectory \
IRVideo folder where the program is installed.
Capture saved: sets the directory for captured images.
Frame rate: determines the number of recorded frames per second (default 25)
Default Palette: sets the palette used in video mode.
Kinescope rate: determines which frame of the image is stored (e.g., 1 - each, 5 - one in
five). In some situations there is no need to record all frames - it helps to save disk space.
Update path view: data refreshing rate for each point.
5.4 Recording and playback of IR videos
5.4.1
Infrared video in real-time mode
Command "Firewire Video..." enables the user to download video directly from the
camera:
The camera must be connected to a computer using USB cable (supplied) through the
USB Interface (11). Camera should operate in real-time mode, no menu/command may
be displayed.
REMARKS:
- Before connecting the camera, it must be switched on and pass self-checks, it is
recommended to connect the camera to a computer before running the program.
- When the camera operates in video mode, it can not use the SD memory card - before
starting the work in video mode and connecting the camera to the computer, remove the