KT-160 USER MANUAL
113
7 Technical specifications
Imaging Characteristics
THERMAL
Detector Type
Uncooled FPA microbolometer (160× 120 pixels, 25
µ
m)
Spectral Range
8-14
µ
m
Thermal sensitivity
≤
0.1°C at 30°C
Field of view /
/ focal length
20.6° X 15.5° / 11mm (standard lens)
Focus
(focus)
Automatic or manual
Screen refresh rate
50/60Hz
VISUAL
Recording visual
image
CMOS sensor, 1600x1200 pixels, "true colours" mode (24-bit)
Image Presentation
External
display
3.6" TFT LCD, resolution 640*480
Video Output
PAL / NTSC
Technology
InfraFusion
Visual and IR blending
Measurement
Temperature range
-20°C up to 250°C (KT-160); -20°C up to 350°C (KT-1 60A);
Accuracy *
± 2°C or ± 2% of reading
Emissivity Correction
Variable from 0.01 to 1.00 (in 0.01 increments)
Measurement
Features
Automatic correction based on distance, relative humidity,
atmospheric transmission and external optics
Optics Transmission
Correction
Automatic based on signals from sensors
*
For non-standard lenses, the manufacturer guarantees the measurement error to be
within specified values only in the central point of the image.