KT-160 and KT-160A USER MANUAL
69
measurement, the FPA detector was calibrated in different temperature points: 7°,
17°, 27°, 37°. For example, when the camera is turn ed on at 0 °, after some time the
temperature of the camera (detector) will gradually increase and when its internal
temperature exceeds 7°, an automatic FPA will take place and will last for about 30
sec. During the FPA setting the camera will not response to any operations.
This action will be repeated when the temperature exceeds 17 ° camera, etc.
In addition, when in use the camera performs a 5-second auto-calibration which is
signalled by the
symbol in the upper left-hand corner of the display.
NOTE!
Use only standard and additional accessories,
listed in "Equipment" section. Using other accessories does not ensure
proper operation of the camera and may cause its damage.
NOTE!
Due to the continuous development of the device, the design
of the display and its certain features may be slightly different than presented
in this manual.
NOTE!
KT-160 and KT-160A Thermal cameras have no parts that could
be repaired by its user. Never attempt
to dismantle or modify the camera. Opening the unit
invalidates the warranty.
NOTE!
Laser locator installed in the camera may be
dangerous to eyes, in case of
direct contact!
DO NOT DIRECTED THE LASER BEAM TOWARDS OTHER PERSONS OR
ANIMALS!
Please note that the laser beam may reflect off
shiny surfaces.
NOTE!
To maintain required parameters of rechargeable batteries, unused
rechargeable batteries should be charged every 3 months. If the device is not
used for a longer period of time, the batteries must be removed from it and
stored separately.