KT-160 USER MANUAL
84
3.8.4
Setting the distance from an object
Select "Distance" and accept it by pressing "OK" to go to sub-menu "Setting the distance
from an object." By pressing the "+" or "-" (buttons
/
) the user may set the
distance between the camera and measured object within the range 0.1 ... 10 meters
(30m when using telephoto lens). Changes are confirmed by pressing "OK" (
).
Setting the distance helps to correct the effects of the medium (air pollution, haze, smoke,
gases), in which the radiation propagates, on the accuracy of the measurements.
The default value is 1.8 m.
3.8.5
Setting relative humidity
Select "Relhum " and accept it by pressing "OK" to go to sub-menu "Setting ambient
relative humidity." By pressing the "+" or "-" (buttons
/
) /) the user may set the
value of relative humidity depending on environmental conditions in the range 0 .. 100%..
Changes are confirmed by pressing "OK" (
).
The default value is 70%.
3.8.6
Activating the laser locator
"Laser" option allows the user to block or make available the possibility of switching on the
laser locator. After entering sub-menu "Switching the laser locator on/off" the user may
activate or block the laser locater by using buttons
/
(current status of the
locator is displayed above the menu bar). After selection for an additional information is
displayed for two seconds in the upper left corner indicating the current status of the laser.
By default, the laser is turned off.
3.9 Camera Settings
When the screen is not showing any menu, press button
and the following function
selection will be displayed: "Parameter - File - Setup".