KT-560M
– USER MANUAL
6
FPA calibration (FPA - Focal Plane Array): in order to ensure accurate temperature
measurement, the FPA detector was calibrated in different temperature points. After
switching the imager on, it performs auto calibration procedure every once in a while.
During the calibration, for about 1 second,
the device does not respond to user’s
activity
, during this time the sound of a shutter clicking twice is heard.
Calibration can be performed manually at any time. In order to do that, the 'C'(2) but-
ton needs to be programmed into 'Calibration' mode (the 'C' button programming pro-
cess is described in chapter 11.1). Then, while in real time mode press 'C'(2) button
and the calibration procedure will be carried out..
During imager operation its housing temperature increases and it’s a normal phenom-
enon.
NOTE!
The thermal imager has no parts that could
be repaired or modified by its user. Never attempt to dismantle or modify the
device. Opening up the unit invalidates the warranty.
NOTE!
Laser locator installed in the camera may be
dangerous to eyes, in case of
direct contact!
DO NOT DIRECTED THE LASER BEAM TOWARDS OTHER PERSONS OR
ANIMALS!
Please note that the laser beam may reflect off
shiny surfaces.
AFTER TURNING THE CAMERA ON, IT PERFORMS INTERNAL TEST, DURING
WHICH, FOR FEW SECONDS, LASER POINTER IS BEING TURNED ON AS
WELL. AFTER TURNING THE CAMERA ON, UNTIL IT REACHES THE POINT
OF BEING FULLY OPERATIONAL IT IS FORBIDDEN TO AIM IT AT HUMAN
AND ANIMALS!!
NOTE!
Use only standard and additional accessories,
listed in "Equipment" section. Using other accessories does not ensure
proper operation of the camera and may cause its damage.
NOTE!
Due to the continuous development of the device, the design
of the display and its certain features may be slightly different than presented
in this manual.