LMW-100 – USER MANUAL
38
5.9
Memory of measurement results
The device remembers the results of the last 20 performed meas-
urements. The device automatically saves the results in the memo-
ry:
in single mode - always,
in continuous mode - the last result before switching off the
function,
in the area and volume measurement
– final result,
in indirect measurement
– final result.
The recording is not performed in the mode of summing up the re-
sults.
To enter the memory browsing mode, press
.
The results can be scrolled using the buttons
.
To clear the memory, in memory browsing mode, press and
hold
and
.
6
Troubleshooting
Error
code
Cause
Solution
204
Calculation error.
Repeat the procedure
208
▪ The beam reflected to the meter is too
weak.
▪ The return time of the beam to the meter
is too long.
▪ Distance to target is >50 m.
Place a shield reflecting the laser beam on
the target.
209
The reflected beam is too strong.
The target is reflecting the beam too hard.
Place a shield reflecting the beam on the
target.
252
Meter temperature is too high.
Cool the meter.
253
Meter temperature is too low.
Warm up the meter.
255
Hardware error.
Switch the meter off and on several times.
If the error persists, contact the service
center.
Summary of Contents for LMW-100
Page 2: ......
Page 3: ...1 INSTRUKCJA OBSŁUGI DALMIERZ LASEROWY MIERNIK WYMIARÓW LMW 100 Wersja 1 00 18 05 2022 ...
Page 8: ...LMW 100 INSTRUKCJA OBSŁUGI 6 3 Opis funkcjonalny 3 1 Wyświetlacz ...
Page 24: ...22 USER MANUAL DISTANCE METER LASER DIMENSION METER LMW 100 Version 1 00 18 05 2022 ...
Page 29: ...LMW 100 USER MANUAL 27 3 Functional description 3 1 Display ...
Page 45: ...43 MANUAL DE USO TELÉMETRO MEDIDOR DE DIMENSIONES LÁSER LMW 100 Versión 1 00 18 05 2022 ...
Page 50: ...LMW 100 MANUAL DE USO 48 3 Descripción funcional 3 1 Pantalla ...
Page 70: ...LMW 100 РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ 68 3 Описание функций 3 1 Дисплей ...
Page 86: ...NOTATKI NOTES NOTAS ПРИМЕЧАНИЯ ...
Page 87: ......
Page 88: ......