MIC-05s1
– USER MANUAL
27
The meter is ready for measure-
ment. Value of interference voltage
can be read on the display.
Measurement
Connect the AutoISO-5000 adapter to the tested cable.
Press
START
to start measurement. Firstly, the checking of volt-
ages on particular pairs of wires is performed.
If any of the voltages exceeds allowable voltage, the
symbol “!” is
displayed (e.g. V
N-PE
!), and the measurement ceases.
View of the screen during
measurement.
After the measurement is
completed the results are
displayed.