MIC-05s1
– USER MANUAL
28
Use
F1
and
F2
buttons to change the group
of displayed results.
Note:
- Remarks and messages are the same as in 3.2.3.
3.2.4 Measurements with increasing voltage - SV
In this mode the meter performs a series of 5 measurements with increasing voltage in steps. The
voltage step-change depends on the set maximum voltage as follows:
- 1 kV: 200 V, 400 V, 600 V, 800 V and 1000 V,
- 2.5 kV: 500 V, 1 kV, 1.5 kV, 2 kV and 2.5 kV,
- 5 kV: 1 kV, 2 kV, 3 kV, 4 kV and 5 kV,
The results for each of the 5 measurements is saved and is signalled by a beep and an icon.
Set the rotary switch of function selection in
the
SV
position. The meter is in the voltage
measurement mode.
Use
F1
,
F2
and
F3
buttons to adjust de-
sired test voltage, capacity of the
object and measurement time, simi-
larly as in section 3.2.2
Press and hold the
START
button for
5 seconds
.
After 5 seconds, the measurement
will be started
and
continued until
the programmed time is reached
or
ESC
button pressed.