MIC-05s1
– USER MANUAL
30
1 minute after closing the circuit the current is measured. The DD value is calculated using the
formula:
C
V
I
DD
pr
min
1
where:
I
1min
– current measured 1 minute after closing the circuit [nA],
V
pr
– test voltage [V],
C
– capacitance [
F].
Set the rotary function selection switch at
DD
. The meter is in the voltage measure-
ment mode.
Use
F1
,
F2
and
F3
buttons to adjust de-
sired test voltage, capacitance of
the object and measurement time
(1 min to 60 min), similarly as in
section 3.2.2
or
Start the measurement similarly as in sec-
tion 3.2.4.
Either during, or after the measure-
ment, the user can switch between
the text display of the results and
the chart by pressing
F4
.