MIC-05s1
– USER MANUAL
36
Press
START
.
An
AUTOZERO
message appears
Confirming that calibration of the test
leads has been performed. The meter
then enters the measurement mode.
The
AUTOZERO
message is displayed
during measurements.
The compensation values are retained
even after the meter is switched off and
on again.
In order to remove the calibration values made above (return to default calibration)
perform the above step with test leads not connected together.
3.4 Temperature measurement
To include temperature measurements for each test function simply connect the temperature
probe. The display indicates the measured temperature along with the measurement results. If the tem-
perature probe is not connected the display shows "T =---". Values are refreshed every second.
To ensure user safety, it is not allowed to mount ST-1 temperature probe on objects with
voltage higher than 50 V to earth. It is advisable to ground the examined object before mount-
ing the probe.