MIC-05s1
– USER MANUAL
61
Measurement of capacitance
Display range
Resolution
Basic uncertainty
0 nF
to
999 nF
1 nF
(5% m.v. + 5 digits)
1.00 µF
to
49.99 µF
0.01 µF
Measurement of capacitance is available only during R
ISO
measurement (when discharging the
object).
Base uncertainty of measurement is met for the tested capacitance connected in parallel with a
resista
nce greater than 10 MΩ.
For measurement voltages below 100 V the measurement error is not specified.
The cable lenght L is calculated as C/Cx, measurement uncertainty depends on measurement
range.
Time Constant TC is calculated as Riso x C, measurement uncertainty depends on measure-
ment range.
Measurement of continuity of protective conductors and equipotential bonding with
200 mA
current
Measuring range according to IEC 61557-4: 0.12
…999
Display range
Resolution
Basic uncertainty
0.00
to
19.99
0.01
±(2% m.v. + 3 digits)
20.0
to 199.9
0.1
200
to
999
1
±(4% m.v. + 3 digits)
Voltage at open terminals: 4 V
to 24 V
Output current at R<15
: min. 200 mA (I
SC
: 200 mA
to 250 mA)
Current flowing bi-directionally, average resistance is displayed on the screen,
Compensation of test leads resistance, autozeroing.
Temperature Measurement
Display range
Resolution
Basic uncertainty
-40.0
to
99.9
°
C
0.1
°
C
±(3% m.v. + 8 digits)
-40.0
to
211.8
°
F
0.1
°
F
±(3% m.v. + 16 digits)
Other technical data
a) insulation .................................................. double-insulated, EN 61010-1 and IEC 61557 compliant
b) measurement category .............................CAT IV 600 V (CAT III 1000 V) according to EN 61010-1
c) degree of housing protection acc. to EN 60529.............................. IP40 (IP67 for closed enclosure)
d) power supply of the meter............................ 14.8 V 5.3 Ah Li-Ion battery, 90 V - 260 V 50 Hz/60 Hz
e) dimensions ................................................ 390 mm x 308 mm x 172 mm / 15.4 in x 12.1 in x 6.8 in
f) meter weight ............................................................................................... approx. 12.3 lbs / 5.6 kg
g) storage temperature ...................................................................... -12
F to 158
F /
–25
C to +70
C
h) working temperature ....................................................................... -4
F to 122
F / -20
C to +50
C
i)
humidity ........................................................................................................................ 20% to 90%
j)
altitude (above sea level) ...................................................................................................
≤3000 m
k) reference temperature ........................................................................... 73
F ± 1.6
F / +23
C ± 2
C
l)
reference humidity ........................................................................................................ 40% to 60%
m) display ............................................................................................................................... TFT LCD
n) number of R
ISO
measurements, acc. to EN 61557-2 with battery power supply ................. min. 1000
o) memory of measurement results .................................................. 990 cells (10 000 records / 8 MB)
p) design, construction and manufacturing are ISO 9001, ISO 14001, PN-N-18001 compliant
q) the device meets the requirements of the EN 61010-1 and IEC 61557 standards
r) the product meets EMC requirements (immunity for industrial environment) according to the
following standards .......................................................................... EN 61326-1 and EN 61326-2-2