10.2 Additional data
Data on additional uncertainties are useful mainly when the meter is used in non-standard condi-
tions and for metrological laboratories for the purpose of calibration.
10.2.1 Additional uncertainties according to EN 61557-2 (R
ISO
)
Significant parameter
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
1% (
BAT
is not lit)
Temperature
32°F to 95°F /
0
C to
35°C
E
3
6%
10.2.2 Additional uncertainties according to EN 61557-4 (R
CONT
)
Significant parameter
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0.2% (
BAT
is not lit)
Temperature
32°F to 95°F
0
°C to 35°C
E
3
1%
11 Accessories
The current list of accessories can be found on the manufacturer's website.
11.1
Standard accessories
The standard set of equipment and accessories supplied includes:
MIC-05s1 meter
set of test leads:
10 kV cable, 3 m, (cat. IV 1000 V), terminated with banana plugs, red
10 kV cable, 3 m, shielded, (cat. IV 1000 V), terminated with banana plugs, black
"E" cable 10 kV, 3 m, (cat. IV 1000 V), terminated with banana plugs, blue
crocodile clip 11 kV DC (cat. IV 1000 V)
– 3 pcs. (black, red and blue)
USB cable
power supply 230V cable
case L-4 for accessories
user manual
calibration certificate
ATTENTION!
11 kV DC crocodile clips are NOT designed to work only on live or energized objects.