The same method applies when measuring the resistance between two conductors of the cable; attach-
ing the foil to the
G
terminal to ignore other conductors that do not take part in the measurement.
For the measurement of insulation resistance of a high voltage circuit breaker, the
G
socket of the
meter should be connected to the insulators of the breaker: