OPERATING MANUAL MIC-10s1 version 1.03
31
After the measurement is
completed the result is displayed.
Press
F4
to display a chart of the measured
resistance and current as a function of time.
Use
,
to move the vertical dashed
cursor line to display measurements at a
selected time. The display will show data
for measured values at the point where the
cursor is positioned. The measurement re-
sult indicates the status of the insulation.
Refer to the following table:
DD value
Insulation condition
>7
Bad
4-7
Weak
2-4
Not the best
<2
OK
Note:
- In environments with strong interference the measurement may be affected by additional uncertainty.
- In the Dielectric Discharge mode test it is also possible to perform the measurement with AutoISO-
5000 adapter. The display of results is similar to the measurement of R
ISO
with AutoISO-5000. The
dsipaly will show the following: