OPERATING MANUAL MIC-10s1 version 1.03
61
10.2 Additional data
Data on additional uncertainties are useful mainly when the meter is used in non-standard condi-
tions and for metrological laboratories for the purpose of calibration.
10.2.1 Additional uncertainties according to EN 61557-2 (R
ISO
)
Significant parameter
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
1 % (
BAT
is not lit)
Temperature
32°F to 95°F /
0
C to
35°C
E
3
6 %
10.2.2 Additional uncertainties according to EN 61557-4 (R
CONT
)
Significant parameter
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0.2% (
BAT
is not lit)
Temperature
32°F to 95°F
0
°C to 35°C
E
3
1 %