OPERATING MANUAL MIC-2501 version 1.04
17
4.3 Voltage measurement
Use
<<
or
>>
button to start the meas-
urement of
(LED
is on). The
meter is in the voltage measurement
mode.
Connect the meter to a
voltage source.
Measurement is per-
formed in a continu-
ous manner.
Additional information displayed by the meter
>750V
, LED is
blinking in red,
two-tone acoustic
signal is generated
Measuring range is exceeded. Voltage is higher than
acceptable.
Immediately
disconnect the test leads.
~
-
When AC voltage is detected, the device will display
symbol "
~
"
("wave") and when DC voltage is detected,
the device will display symbol "
-
" for negative polarity
or "nil" for positive polarity.