f) meter weight .................................................................................................... approximately 1.3 kg
g) storage temperature ..................................................................................................... –20...+70
°
C
h) working temperature ...................................................................................................... -10...+40
°
C
i) nominal temperature ........................................................................................................ +23 ± 2
°
C
j) number of R
ISO
measurements ........................................................................... approximately 900
k) display ........................................................................................................................ LCD segment
l) quality standard ........................... development, design and manufacturing are ISO 9001 compliant
m) the device meets the requirements of IEC 61557 standard
n) the product meets EMC requirements (immunity for industrial environment) according to the follo-
wing standards .............................................................. EN 61326-1:2006 and EN 61326-2-2:2006
9.2 Additional data according to IEC 61557-2 (R
ISO
)
Data on additional uncertainties are useful mainly when the meter is used in non-standard condi-
tions and for metrological laboratories for the purpose of calibration.
Significant parameter
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0% (
BAT
is not lit)
Temperature 0...35°C
E
3
0.1%/ºC
10 Equipment
10.1
Standard equipment
Standard set of equipment supplied by the manufacturer includes:
•
mater MIC-2505 –
WMPLMIC2505
•
set of test leads:
5kV leads, 1.8m, cat. IV 1000V terminated with banana plugs - 2 pcs. (red –
WAPRZ1X8REBB
and blue –
WAPRZ1X8BUBB
)
5kV shielded lead, 1.8m, cat. IV 1000V, terminated with banana plugs - ( black –
WAPRZ1X8BLBB
•
5kV crocodile clip, cat. IV 1000V – 2 pcs. (Black –
WAKROBL20K04
and blue –
WAKROBU20K05
)
•
blade probe with banana socket – 1 pcs. (red –
WASONREOGB2
)
•
battery pack –
WAAKU10
•
adapter for battery pack charging –
WAZASZ7
•
CD SONEL
•
meter harness –
WAPOZSZE2
•
carrying case
L-4 –
WAFUTL4
•
operating manual
•
calibration certificate