OPERATING MANUAL MMR-630 Revision 1.6
15
3.1.2 Flow of current
Measurements may be performed with a single-direction current or bidirectional current flowing in
opposite directions. In the latter case, the principal result which is displayed is the average resistance
value. Measurements with single-direction current are faster in the case of objects without internal
tension and electrothermal forces. On the other hand, measurements by means of bidirectional cur-
rent eliminates errors resulting from the presence of such tensions and forces in the measured object.
If the single-direction flow of current is selected the resistance value R
R
is not displayed.
3.1.3 Measurement triggering
The MMR-630 meter has three modes of measurement triggering: normal, automatic and contin-
uous.
In the normal mode one measurement of resistance is performed.
In the automatic mode the device waits for all the four test leads to be connected to the object,
and then the measurement is automatically started.
In the continuous mode for resistance objects the meter performs subsequent measurement cy-
cles: measurement of resistance and display of the result for approximately three seconds.
In the continuous mode for inductive objects the meter performs one measurement cycle with a
single-direction current (RF). During the cycle the result is gradually stabilised, and it is displayed
approximately each half a second.
Summary of Contents for MMR-630
Page 1: ......
Page 2: ......
Page 3: ......
Page 12: ...OPERATING MANUAL MMR 630 Revision 1 6 9...
Page 27: ...OPERATING MANUAL MMR 630 Revision 1 6 24 4 3 Memory browsing 4 4 Memory erasing...
Page 28: ...OPERATING MANUAL MMR 630 Revision 1 6 25...
Page 38: ...OPERATING MANUAL MMR 630 Revision 1 6 35...
Page 39: ......
Page 40: ......