Notes:
Continuous triggering for inductive object may be used solely when the meter is
permanently connected to the power grid. It is unacceptable to disconnect the me-
ter from the object during measurements. This is dangerous for the user and may
cause damage to the device.
- In the case of measurements of inductive objects the automatic measurement triggering mode is not
active.
3.1.4 Saving to memory
Automatic saving of the measurement results to memory may be particularly useful in the automat-
ic and continuous measurement triggering modes. Saving is performed directly after the result has
been displayed in the screen.
3.1.5 Measurement range selection
For the purpose of measurement of the resistance of the object the MMR-630 meter applies
measurement currents within the range between 0.1mA and10A. The measurement range and thus
the measurement current may be selected automatically or manually.