MMR-650 – USER MANUAL
26
Auto/Manual switch (auto /
manual range selection):
Auto - the measuring current
is limited to a value not ex-
ceeding the set value,
Manual - measuring current
has the set value.
Setting the measuring current.
Use
button to go to the next settings.
Switching Unidirectional/Bidi-
rectional measurement.
Selecting trigger method:
- manual: use
START/STOP
button
- automatic: by connecting test
leads to the object
- continuous: the measurement
is triggered and ended by
pressing
START/STOP
button.
Boxes for setting the range and measurement current on the screen above correspond to the same
boxes shown in the previous screen.
Note:
- Bidirectional measurement is used to compensate the potential present at the contact point of two
different conductors. In this case, the main result of the measurement is the average of results from
individual directions.
- After selecting automatic trigger, the first measurement must be started by pressing
START/STOP
button, next measurements may be initiated by connecting the test leads to the object.
- Automatic trigger operates correctly for resistance
≤
4 k
Ω
, above this value there is no guarantee of
correct operation.