MMR-650 – USER MANUAL
33
Incorrect result" R
PE
outside limits.
R
F
– resistance at the measuring current flowing in the assumed positive direction
R
R
– resistance at the measuring current flowing in the assumed negative direction
T
a
– ambient temperature
T
1
– object temperature
T
0
– reference temperature
α
- temperature coefficient of resistance
- upper limit
- lower limit
Use
button to display the
graph of resistance over time.
Note:
- It is recommended to use the maximum source power (set the current limit high enough), because then
the core is saturated faster and the result stabilizes faster.
- Declared measurement uncertainty refers to the measured value; for value calculated with the refer-
ence temperature the uncertainty is not specified.
- Resistance values> 2 k
Ω
, displayed on the screen during the measurement are shown for information
purposes - they have no specified accuracy.
- The result may be entered into the memory by pressing
button (see sec. 4.2).