MMR-650 – USER MANUAL
35
Switching between Auto/Man-
ual mode:
Auto - the measuring current
is limited to a value not ex-
ceeding the set value,
Manual - measuring current
has the set value.
When measuring inductive ob-
jects, only Auto mode is avail-
able.
Setting the measuring current
values or limits.
Use
button to go to the next settings.
Selecting the
object type:
- resistance
- inductive
Selecting the time in-
terval: 1 s, 5 s, 10 s,
15 s, 1 min, 5 min, 10
min, 15 min.
Selecting logging
time: 1 min, 5 min, 10
min, 15 min, 30 min.
Note:
- The selected sampling time must be lower than the time set for logging, otherwise the meter will set
the default times.
Select resistance limits from the database (click
Limit settings’ name
) or set the required
limits for the performed measurement by clicking
Lower limit
and/or
Upper limit
.