10 Technical data
10.1 Basic data
abbreviation
„w.m.” used in the specification of basic uncertainty signifies standard measured value
Measurement of alternating voltages (True RMS)
Range
Resolution
Basic uncertainty
0.0...299.9 V
0.1 V
(2% w.m. + 6 digits)
300…500 V
1 V
(2% w.m. + 2 digits)
Frequency range: 45...65 Hz
Measurement of frequency
Range
Resolution
Basic uncertainty
45.0...65.0 Hz
0.1 Hz
(0.1% w.m. + 1 digit)
Voltage range: 50...500 V
Measurement of short circuit loop impedance Z
L-PE
, Z
L-N
, Z
L-L
Measurement of short circuit loop impedance Z
S
Test range according to IEC 61557:
Test lead
Test range Z
S
1.2 m
0.13...1999
5 m
0.17...1999
10 m
0.21...1999
20 m
0.29...1999
WS-03, WS-04
0.19...1999
Display range:
Display range
Resolution
Basic uncertainty
0...19.99
0.01
(5% w.m. + 3 digits)
20.0...199.9
0.1
200...1999
1
Nominal working voltage U
nL-N
/ U
nL-L
: 110/190 V, 115/200 V, 127/220 V, 220/380 V, 230/400 V,
240/415 V
Working range of voltage
: 95…270 V (for Z
L-PE
and Z
L-N
) and
95…440 V (for Z
L-L
)
Nominal network frequency f
n
: 50 Hz, 60 Hz
Working range of frequency
: 45…65 Hz
Maximum test current (for 415 V): 41.5 A (10 ms)
Test of correct PE terminal connection with a touch electrode
Indications of short circuit loop resistance R
S
and short circuit loop reactance X
S
Display range
Resolution
Basic uncertainty
0..19.99
0.01
(5% + 5 digits) of Z
S
value
Calculated and displayed for a value of Z
S
<20