MPI-535
– BRIEF USER MANUAL
39
Discharging the object tested
Measurement of insulation resistance with the use of UNI-Schuko plug (WS-03, WS-04)
between all three terminals (for U
N
=1000 V is not available)
Insulation resistance measurement for multi-wire cables (max. 5) using an optional exter-
nal AutoISO-1000c adapter
Measurement of voltage on terR
ISO
, -R
ISO
within the range of: 0 V...440 V
Test current < 2 mA
7.1.9
Light measurements
Measuring ranges of LP-1 probe
Range
[lx]
Resolution
[lx]
Spectral
uncertainty
Basic
uncertainty
0...399.9
0.1
f1<6%
(5% m.v. + 5 digits)
400...3999
1
4.00 k...19.99 k
0.01 k
Probe class B
Measuring ranges of LP-10B probe
Range
[lx]
Resolution
[lx]
Spectral
uncertainty
Basic
uncertainty
0...39.99
0.01
f1<6%
(5% m.v. + 5 digits)
40.0...399.9
0.1
400...3999
1
4.00 k…39.99 k
0.01 k
40.0 k…399.9 k
0.1 k
Probe class B
Measuring ranges of LP-10A probe
Range
[lx]
Resolution
[lx]
Spectral
uncertainty
Basic
uncertainty
0...3.999
0.001
f1<2%
(2% m.v. + 5 digits)
4.00...39.99
0.01
40.0...399.9
0.1
400...3999
1
4.00 k…39.99 k
0.01 k
40.0 k…399.9 k
0.1 k
Probe class A
7.1.10 Phase sequence
Phase sequence indication: in the same direction (correct), opposite direction (incorrect)
Range of network voltages U
L-L
: 95 V.
..500 V (45 Hz…65 Hz)
Display of phase-to-phase voltages