OPERATING MANUAL MPU-1 version 1.5
15
After pressing
ENTER
the device automatically displays "Alarm on" and begins to monitor current.
The most recent threshold setting is retained even after it is turned OFF and ON again.
CAUTION:
In "Alarm on" mode - the display turns off after 5 min to reduce energy drain on the battery.
Pressing any button will re-activate the display.
6.4.2 Displaying and Measuring Leakage Current in "Alarm off" mode.
The MPU-1 can be used to measure and display leakage current without the monitoring and alarm
function; “Alarm off mode”. (Note that whenever the MPU-1 is turned on it is always first activated in
“Alarm on” mode.)
To measure and display the leakage current in
“Alarm off” mode:
Connect probes to the socket:
"Probe (1)" 7
to perform single-probe measurement; or to sockets
"Probe (1)" 7
and
"Probe (2)" 8
to perform dual-probe measurement
.
Press
MENU
button.
,
Use
,
buttons to select
Alarm
.
Confirm your choice by pressing
ENTER
button .