Additional information displayed by the meter
The meter is ready for measurement.
Voltage on terminals
L
and
PE
is not within the
measurable range.
L and N conductors have been switched (voltage between
terminals
PE
and
N
).
The temperature inside the meter has risen above the
limit, the measurement is blocked.
RCD is inactive.
Safe contact voltage exceeded.
R
E
value is out of range.
The measurement cancelled with
ESC
.
The loss of voltage during the measurement.
After U
B
R
E
measurement, I
A
(or t
A
) measurement has not
been performed because the values of R
E
and mains
voltage did not allow to generate the required current
value.
or
Damaged current setting circuit. Try to perform
measurement again. If the message reappears, please
send the meter for repair.
2.6.2 Measurement of RCD disconnection time
Turn on the meter.
Turn the rotary switch
to a position for
t
A
measurement with
selected multiplication
factor of I
Δ
n
.
Set the parameters according to the following algorithm, and
according to the rules described in general parameters setting.