MRU-120
– USER MANUAL
8
Once the calibration has been successfully concluded,
The following will be displayed.
The meter has determined the correction factor for connected clamp. The factor is saved in the
memory also when the power supply of the meter is off until the following successful calibration of the
clamp has been performed.
Notes:
- Make sure the test lead passes centrally through the clamp.
Additional information displayed by the meter
Message
Cause
Procedure
ERROR: CLAMP NOT CONNECTED
OR NOT PUT ON WIRE
CONNECTED TO H AND E SOCKET!
CALIBRATION ABORTED.
PRESS
ENTER
The clamp is not
connected
Check whether the
clamp is connected
to the device or
whether it is placed
upon the test lead
used by the meter
to force the pas-
sage of current.
ERROR: WIRE NOT CONNECTED
TO H AND E TERMINAL!
CALIBRATION ABORTED.
PRESS
ENTER
No wire
Revise the connec-
tions
ERROR: CALIBRATION
COEFFICIENT OUT OF RANGE.
CALIBRATION ABORTED.
PRESS
ENTER
Incorrect calibra-
tion factor
Check the quality of
the connections
and/or replace the
clamp.