MRU-120
– USER MANUAL
33
Note:
The displayed
BAT
symbol means insufficient power supply voltage and the need to charge the
accumulators,
Measurements realized with an insufficient meter power supply voltage are distorted with addi-
tional errors which are impossible to ascertain by the user and thus they cannot constitute a basis
for a conclusion of correctness of the tested earthing system.
6.2 Replacement of accumulators
The MRU-120 meter is equipped with a package of NiMH accumulators and charger.
The package of accumulators is placed in a compartment. The charger is installed inside the meter
casing and it may be used solely to charge the original accumulators. It is powered from an external
power supply. It is also possible to use a car lighter socket.
WARNING:
If the test leads are left in the sockets during replacement of the batteries or the
package of accumulators, there is a risk of electric shock with a dangerous
voltage.
In order to replace the package of accumulators it is necessary to do the following:
Remove all the test leads from the sockets and turn the meter off,
Remove the four screws of the accumulators/batteries compartment (in the lower part of the
casing),
Remove the compartment,
Insert the compartment in the meter,
Replace the four screws of the accumulators/batteries compartment.
NOTE!
Do not use the meter when the accumulator compartment is removed or open
or power it from other sources than those mentioned in the present manual.