MRU-120
– USER MANUAL
38
10 Technical data
The abbreviation „m.v.” in the accuracy definition means the measured value.
10.1 Basic data
Interference voltage measurement U
N
(RMS)
Range
Resolution
Accuracy
0...100 V
1 V
±(2% m.v. + 3 digits)
measurement for f
N
4
5…65 Hz
frequency of measurements
– minimum two measurements/s
Measurement of resistance of protective conductors and equipotential bonding (2P)
The measurement method: in accordance with IEC 61557-4
Range of measurement in accordance with IEC 61557-4: 0,24
... 19,9 k
Range
Resolution
Accuracy
0,00...19,99
Ω
0,01
Ω
±(2% m.v. + 2 digits)
20,0…199,9 Ω
0,1
Ω
200…1999 Ω
1
Ω
2,00…9,99 kΩ
0,01
Ω
±(5% m.v. + 2 digits)
10,0…19,9 kΩ
0,1
kΩ
Measurement of earth resistance
– 3-pole method (R
E
3P), 4-wire method (R
E
4P)
The measurement method: 3-pole, in accordance with IEC 61557-5
Range of measurement in accordance with IEC 61557-5: 0,30
... 19,9 k
Range
Resolution
Accuracy
0,00...19,99
Ω
0,01
Ω
±(2% m.v. + 2 digits)
20,0…199,9 Ω
0,1
Ω
200…1999 Ω
1
Ω
2,00…9,99 kΩ
0,01
kΩ
±(5% m.v. + 4 digits)
10,0…19,9 kΩ
0,1
kΩ
Measurement of resistance of auxiliary electrodes R
H
and R
S
Range
Resolution
Accuracy
0...999
Ω
1
Ω
±(5% (R
E
+R
H
+R
S
) +
8 digits) but not less
than 10% R
E
1,00...9,99
kΩ
0,01
kΩ
10,0…19,9 kΩ
0,1
kΩ