MRU-200-GPS OPERATING MANUAL Version 1.06
9
The meter has determined the calibration coefficients for the connected clamp or probe. This da-
ta is saved in the memory and is retained when the meter is switched off. In any subsequent calibra-
tion of a clamp or probe the calibration data will be updated.
Calibration of flexible clamps (using ERP-1 adapter)
After reading the intro-
ductory information
press
ENTER
.
Follow on-screen prompts displayed by the meter and short H and E sockets with
a wire.
Connect ERP-1 adapter to the terminal of the clamps.
Turn ERP-1 adapter ON.
Connect flexible clamps to ERP-1 adapter.