MRU-200-GPS OPERATING MANUAL Version 1.06
40
Use buttons and to set the dis-
tance between probes and press
ENTER
to begin measurement.
View the result.
Current probe resistance
Voltage probe resistance
Additional uncertainty caused by
the resistance of the probes
By pressing the
F4
button you can display GPS coordinates.
The result is displayed for 20s. It may be displayed again when
ENTER
is pressed.
Notes:
Earth ground resistance measurement may be performed successfully if the in-
terference voltage does not exceed 24V. The interference voltage is measured
up to 100V, but over 50V is it signalled as hazardous. Do not connect the meter
to a voltage exceeding 100V.