PQM-700 Operating Manual
74
7.13 Safety and electromagnetic compatibility
Safety and EMC
Compliance with
IEC 61010-1, 3
rd
Edition
Measurement Category
(Voltage measurement inputs)
IV 300 V
III 600 V
II 760 V
pollution class 2
Overvoltage Category
(AC/DC Power adapter)
IV 300 V
III 415 V
III 460 V (including fluctuations)
pollution class 2
Insulation
Double acc. to IEC 61010-1
Electromagnetic compatibility
IEC 61326
Immunity to radio frequency interferences
IEC 61000-4-3
sinusoidal modulation 80% AM, 1kHz
80…1000MHz, 10V/m
1.4…2.0 GHz, 3 V/m
2.0 ... 2.7 GHz, 1 V/m
Immunity to electrostatic discharge
IEC 61000-4-2
Air discharge: 8 kV
Contact discharge: 4kV
Immunity to conducted disturbances, in-
duced by radio-frequency fields
IEC 61000-4-6
sinusoidal modulation 80% AM, 1kHz
0.15…80MHz, 10V
Immunity to a series of electrical fast tran-
sients/bursts
IEC 61000-4-4
Amplitude of 2kV, 5kHz
Surge immunity
IEC 61000-4-5
Amplitude 2kV (L-L)
Emission of radiated RF disturbances
IEC 61000-6-3
30…230MHz, 30dB(
V/m) at 10m
230…1000MHz, 37dB(
V/m) at 10m
Emissions of conducted interferences
IEC 61000-6-3
Levels for a quasi-peak detector:
0.15kHz…0.5MHz: 66dB
V…56dB
V
0.5MHz…5MHz: 56dB
V
5MHz…30MHz: 60dB
V
7.14 Standards
Standards
Measurement methods
IEC 61000-4-30 Class S
Measurement accuracy
IEC 61000-4-30 Class S
Power Quality
EN 50160
Flicker
IEC 61000-4-15
Harmonics
IEC 61000-4-7
Safety
IEC 61010
EMC
IEC 61326
Quality standard
design, construction and manufacturing are ISO 9001 compliant
Summary of Contents for PQM-700
Page 85: ...85 Notes...