GENERAL INFORMATION 

 

The ultrasonic power supply converts 50/60 Hz line voltage to high frequency electrical 

energy at 20 kHz or 40 kHz (20,000 or 40,000 cycles per seconds), depending on the 

model. This high frequency electrical energy is transmitted to the piezoelectric transducer 

within the converter, where it is changed to mechanical vibrations. The ultrasonic 

vibrations are intensified by the probe and focused at the tip where the atomization takes 

place. The liquid travels through the probe, and spreads out as a thin film on the 

atomizing surface. The oscillating tip disintegrates the liquid into micro-droplets, and 

ejects them to form a gentle, low velocity spray. 

 

Unlike conventional atomizing nozzles that rely on pressure and high-velocity motion to 

shear a fluid into small drops, the ultrasonic atomizer uses only low ultrasonic vibrational 

energy for atomization. The liquid can be dispensed to the atomizing probe (nozzle) by 

either gravity or a small low-pressure metering pump, and atomize continuously or 

intermittently. The rate at which the liquid is atomized depends, within limits, solely on 

the volume that is being delivered onto the atomizing surface, and the frequency. The 

amount of material atomized can be as little as 2 µl/sec. Because the droplets typically 

drift downward at low velocity under the influence of gravity, the probe should be 

mounted with tip facing downward, and air disturbances kept to a minimum. A wide 

variety of coatings, chemicals, lubricants, and particulate suspensions can readily be 

atomized. However, factors such as viscosity, miscibility, and solid content deserve 

consideration. For optimum atomization, the viscosity should be under 500 cps and the 

solid concentration kept below 30%. Because the atomization process depends on setting 

a liquid film into motion, typically the higher the viscosity – the lower the flow rate, and 

the more difficult the application. The atomization of liquids containing long-chained 

polymer molecules is problematic, even in diluted form, due to the highly cohesive nature 

of the material. In many cases, mixtures with particulates can be atomized, because the 

solids are simply carried along in the drops. The low transport velocity of the liquid 

through the probe permits even abrasive slurries to be processed with negligible erosion 

of the passageway. Compared with conventional pressurized nozzles, the feed channel 

running through the probe and exit orifice are relatively large, and practically 

uncloggable. Drop size is primarily a function of frequency, and the higher the frequency, 

the smaller the drop diameter. The median drop size at 40 kHz is 50 microns. 

 

Atomizing probes are one-half wavelength long tools that act as mechanical transformers 

to increase the amplitude of vibration generated by the converter. They are fabricated 

from high grade titanium alloy Ti-6Al-4V because of its high tensile strength, good 

acoustical properties at ultrasonic frequencies, high resistance to corrosion, low toxicity, 

and excellent resistance to cavitation erosion.  

 

 

 

 

 

 

Summary of Contents for VCX 130 AT

Page 1: ...ds General Information SECTION 1 INSTALLATION Inspection Electrical Requirements Installing the Ultrasonic Atomizer SECTION II OPERATION Functions of Controls Indicators and Connectors Preparations fo...

Page 2: ...in connection with the sale of its products The manufacturer hereby disclaims any warranty of either merchantability or fitness for a particular purpose No person or company is authorized to change mo...

Page 3: ...in a manner contrary to that specified in this instruction manual the protection features designed into the unit may be impaired When mounting the atomizer in a stand always clamp the converter housi...

Page 4: ...ces kept to a minimum A wide variety of coatings chemicals lubricants and particulate suspensions can readily be atomized However factors such as viscosity miscibility and solid content deserve consid...

Page 5: ...0 60 Hz at 100 volts 115 volts 220 volts or 240 volts depending on the voltage option selected For power requirements check the label on the back of the unit Should it become necessary to convert the...

Page 6: ...free from excessive dust dirt explosive and corrosive fumes and extremes of temperature and humidity CAUTION Operating the equipment with voltages different from those specified or with voltages with...

Page 7: ...REVIEW Key Used to enter the amplitude selected and view amplitude power and energy START STOP Key Starts or stops the ultrasonics In the STOP mode the red indicator goes off Key Switches the main pow...

Page 8: ...s the fuse s 9 PIN D SUB CONNECTOR Pin No Description 1 Not connected 2 Not connected 3 Not connected 4 Enables connection to a frequency counter 5 Enables connection to an external power monitor 5 mv...

Page 9: ...he converter housing only Do not secure the clamp to the probe 5 Connect the converter cable to the power supply CAUTION Do not operate an atomizer that has been in a very cold or hot environment for...

Page 10: ...ination of level sensor and inlet valve Pressure reducing regulators should be used when working with high pressure delivery systems Optimum atomization can only be obtained by empirically adjusting t...

Page 11: ...11 Do not operate the probe when the ambient temperature or temperature of the liquid exceeds 158 F 70 C...

Page 12: ...sonics is off press the AMPL key and the key for a 40 reading on the screen the screen will display Amplitude Setting 40 and then press the ENTER REVIEW key The screen will display The atomizer is now...

Page 13: ...urn the ultrasonics off plug in the footswitch or remote actuation device into the footswitch jack and depress the switch again to turn the ultrasonics back on Depressing the footswitch or closing the...

Page 14: ...s has failed proceed as follows 1 Turn the unit off by depressing the key and disconnect the line cord from the electrical outlet 2 Open the fuse holder cover using a small screwdriver and pull out t...

Page 15: ...necessary delay Care should be exercised to provide adequate packing to insure against possible damage in shipment The atomizer should be sent to the Service Department with all transportation charges...

Reviews: